8MOXI / SelectAir distributor prior to operating the control unit. Do not alter the plug to fitinto the power outlet. The power supply cord should be

Page 3 - Table of Contents

9Section 5 InstallationInstallation - Mattress ReplacementsMattressNote: Make sure that the mattress hose fitting is at the foot end

Page 4

10 4. Place the ROHO HIGH PROFILE cushion into the HealFloat foot section of the mattress overlay. Note: Refer to the cushion's enclosed

Page 5 - Softer Firmer

Section 5 Installation4. Ensure the power entry switch is “OFF”.5. Firmlyinsertthepowercordintotheoutletonthecontrolunit(fig.6).6. In

Page 6 - Figure 1

2. Press the firm mattress timer.3. Place the patient on the mattress, centered and in their habitual resting position, i.e. on their back, on their

Page 7 - Figure 2

Section 6 Manual Setup Comfort Level Reference Chart Note: These charts are only a guideline because weight distribution varies from

Page 8

Note: Before opening, observe what voltage is visible on the red fuse holder through the window on the fuse holder door.4. Carefully remove

Page 9

FilterThe filter should be inspected and cleaned every month under normal conditions. Inspect filter and clean or replace as necessary. Bariatric pati

Page 10

Cleaning and Disinfecting Note: The following instructions are intended only as general guidelines. Follow yourfacility’s standard protocol for clean

Page 11 - Section 5

washing machine should be set at a regular wash, hot water setting.4. Launder the sheet in a normal manner using the regular cycle and hot water

Page 13 - Section 5 Installation

18HealFloat feature (HybridSelect only)Refer to the ROHO HIGH PROFILE cushionʼs operating manual for specific cleaning instructions.Control Unit and H

Page 14 - Figure 10

19TroubleshootingProblem/Corrective Action1. Control unit does not function, and power light not illuminated.Ensure the power cord is connected to a f

Page 15

20Section 9 TroubleshootingCheck hose for crushed or kinked areas, or internal blockages.Remove filter tray and ensure air inlet vent in control unit

Page 16 - Section 7 Maintenance

21SelectAir Upright ModeThe Upright Mode increases system pressure to compensate for the upright patient.Note: When the patient is sitting up in bed,

Page 17

22Note: To assist in deflating the mattress, disconnect the hose at the mattress or control unit.Thenpressthe"UprightOn/Off"

Page 18

23Section 11 SelectAir MaxFigure 17SelectAir Max Upright ModeThe Upright Mode increases system pressure to compensate for the upright patient.Note:

Page 19

24Figure 19Section 11 SelectAir MaxTo deactivate the Upright Mode, first lower the patient to a lying position. Thenpress the "Upright On/Off&quo

Page 20

Figure 21LightSection 11 SelectAir MaxTo unlock the control panel, press and hold the “System Lock” button for three (3)seconds. EnsuretheSystem

Page 21 - Section 9 Troubleshooting

262. The Pulsate Mode adjusts as follows: Low–20% cycle every 15 minutes Medium–20% cycle every 7 minutes High–20% cycle every minute.3. To

Page 22

27Section 12 BariSelectFigure 25BariSelectUpright ModeThe Upright Mode increases system pressure to compensate for the upright patient.Note: When th

Page 23 - Section 10 SelectAir

1ENGLISH Section 1 Introduction 2-3 Section 2 System Components 4-5 Section 3 Safety Precautions 6 Section 4 Important Safeguards 7-8 Sec

Page 24

28To deactivate the Upright Mode, first lower the patient to a lying position. Thenpress the "Upright" button a second time. Ensure the Upr

Page 25 - Section 11 SelectAir Max

29 • To lock the control panel, press the “Lock/Unlock” button. If no patient activity is detected for 60 seconds, the control unit will auto

Page 26 - Section 11 SelectAir Max

30Manual Weight Input BariSelectWheninnormaloperatingmode,presstheManualbuttononthekeypad(fig.30).Thecontrolunitwillentermanualwe

Page 27

31Figure 33Note: To assist in deflating the mattress, disconnect the hose at the mattress or control unit.CPR/Emergency Deflate Importan

Page 28

32HybridSelectUpright ModeThe Upright Mode increases system pressure to compensate for the upright patient.Note: When the patient is sitting up in be

Page 29 - Section 12 BariSelect

33Section 14 HybridSelectTo deactivate the Upright Mode, first lower the patient to a lying position.Then press the "Upright On/Off" button

Page 30 - Section 12 BariSelect

34Light2. Tounlockthecontrolpanel,pressandholdthe“SystemLock”buttonforthree(3) seconds. Ensure the System Lock light extinguishes.

Page 31

35Select Air System Specifications:SelectAirSelectAirMaxBariSelect(all sizes)HybridSelectMattress System TypeMattressreplacementMattressreplacementMatt

Page 32

36Pulsation - Continually readjusts system pressure. SelectAir Max & BariSelect -Low-every 15 minutes, Medium-every 7 minutes, High-every minute.

Page 33

SelectAir 5 Year Product Warranty37Section 15 WarrantyMOXI Enterprises, LLC warrants the SelectAir (Model SASTDPUMP), SelectAirMax (Model SAMAXPUMP),

Page 34 - Section 13 HybridSelect

2IntroductionSelectAir® low air loss mattress systems are pressure relief mattress systems designed for the treatment and/or prevention of pressure ul

Page 35 - Section 14 HybridSelect

1939 South Vandeventer Avenue, Saint Louis, Missouri 63110www.moxiusa.comToll Free: 1.800.942.7678 • Local: 314.771.5222 • Fax: 314.771.5225of North A

Page 36

3Section 1 IntroductionSymbol LegendAttention- See accompanying instructions for useor information found in this manual or labels on the

Page 37

4System Components For Mattress Replacement ModelsSection 2 System ComponentsNote: SelectAir and SelectAir Max mattress replacements

Page 38

5System Components For HybridS elect Mattress Overlay ModelsSection 2 System ComponentsFigure 2MattressOverlayControl Unit with flexible h

Page 39 - Section 15 Warranty

6Safety PrecautionsPlease read and thoroughly understand the following safety precautions before using this product.WARNINGS• Explosion risk if used

Page 40

7Important SafeguardsPlease read and follow the instructions regardingSelectAir mattress and overlay systemsand retain this manual for future referenc